Gauge

572 Bayard-Alpert Ion Gauge Tube

by samhosys posted Apr 17, 2023
?

단축키

Prev이전 문서

Next다음 문서

ESC닫기

크게 작게 위로 아래로 댓글로 가기 인쇄

572.jpg

 

The 572 Bayard-Alpert Gauge hot-cathode ionization gauge tube is a high-performance, high vacuum gauge with a wide range of linear response from 2 x 10-10 to 1 x 10-3 Torr (2.7 x 10-10 to 1 x 10-3 mbar). It is designed to withstand long periods of outgassing, and to facilitate simple external switching of filaments without disturbing the vacuum system. The 572 is compatible with virtually all modern hot filament ionization gauge controllers.

 

 

 

 

 

 

 

 

 

 

 

 

Features


 

-Dual tungsten filament keeps gauge operational even when one filament burns out

 

-Robust instrument that can withstand long periods of degassing and accidental exposure atmosphere at

operating temperature and still recover its orignal characteristics

 

-High X-ray limit allows the gauge to give more accurate readings at lower pressure

 

 

 

 

 

 

 

Specifications


 

Sensitivity 10(Torr)-1(mbar)-1 (typical)
Operating ratings

0 VDC to ground (collector)

+180 VDC to ground (grid)

+30 VDC to ground (filament)

X-Ray limit 2 x 10-10 Torr, (2.6 x 10-10 mbar)
Operating pressure

2 x 10-10 Torr to 1 x 10-3 Torr

(2.7 x 10-10 mbar to 1 x 10-3 mbar)